Walchem WebMaster Manuel d'instructions Page 8

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 180
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 7
4
Chlorine, Chlorine Dioxide sensors
Pressure: 0 to 1 atm (0 to 14.7 psi)
Temperature
Chlorine Dioxide sensors: 41 to 122°F (5 to 50°C)
Chlorine: 41 to 113°F (5 to 45°C)
Flow Cell Inlet: ¼” NPTF
Flow Cell Outlet: ¾” NPTF
Wetted Materials of Construction: PVC, Silicone or PTFE, Nylon, Isoplast, FKM
High pressure tower contacting conductivity electrode
Pressure: 0-300 psi
Temperature: 32-158°F (0 to 70°C)
Process connections: ¾” NPTM
Wetted Materials of Construction: 316 SS, PEEK
High pressure boiler contacting conductivity electrode
Pressure: 0-250 psi
Temperature: 32-401°F (0 to 205°C)
Process connections: ¾” NPTM
Wetted Materials of Construction: 316 SS, PEEK
High pressure pH electrode
Pressure: 0-300 psi
Temperature: 32-275°F (0 to 135°C)
Process connections: ¾” NPTM
Wetted Materials of Construction: Glass, Polymer, PTFE, 316 SS, FKM
High Pressure ORP electrode
Pressure: 0-300 psi
Temperature: 32-275°F (0 to 135°C)
Process connections: ¾” NPTM
Wetted Materials of Construction: Platinum, Polymer, PTFE, 316 SS, FKM
High Pressure flow switch
Pressure: 0-300 psi
Temperature: 32-158°F (0 to 70°C)
Process connections: ¾” NPTF
Wetted Materials of Construction: Carbon steel, Brass, 316SS, FKM”
Vue de la page 7